Read-only Modbus RTU register map and working example clients for the Pisonics PS7000 ultrasonic slurry density meter — a non-nuclear inline density meter that measures acoustic impedance, so it needs no radioactive source, no licence and no disposal route.
Everything here is generated from a single source of truth
(registers/input-registers.json) and the
decoder is unit tested, so the parts that are easy to get wrong are checked
rather than assumed.
- 📋 Register map — the tables
- 🔢
input-registers.json·.csv— machine readable ⚠️ Byte order and the off-by-one — read this first- 🩺 Status and error codes — including how to tell a stale reading
- 🐍 Python example · 🔷 C# example
1. Register numbers here are 1-based; the frame address is one lower. To read register 1026, put 1025 in the request. Get it wrong and the block still reads — just shifted by one, giving a plausible but wrong number.
2. 32-bit values are CDAB — low word first. Most masters default to ABCD. If floats decode as denormals or zeros, this is why.
3. Read the block in one request. The firmware publishes a consistent snapshot; field-by-field reads are slower and can mix measurement cycles.
Quick check that you have both right: read register 805. It must return
0x50 (density / concentration meter) or 0x51 (spectral concentration
meter). Anything else means you are off by one.
cd examples/python
pip install -r requirements.txt
python read_ps7000.py --port /dev/ttyUSB0 --slave 1 density: 1.4210 g/cm3
probe temperature: 31.50 degC
signal quality: 0.930
loop 1: 11.203 mA
up time: 86400 s
entrained gas (EMA): 4 / 100
The decoding lives in ps7000.py, which has no
dependencies at all. You can lift that one file into your own project and
ignore the rest. Its tests run without an instrument:
python test_ps7000.pymainValueFiltered — register 1026, F32. Read mainValueUnit
(register 1028) alongside it, because the same instrument can be configured to
report density in g/cm³ or kg/m³, mass or volume concentration, Baumé, or
solids in g/L. A host that assumes one will be wrong by roughly 100× the day
someone changes it on the panel.
Use measureErrorCode (register 1004), not the ok field (register 1029).
ok reports sampling quality for the cycle. On aerated or turbulent slurry it
is routinely 0 while the value is perfectly live and the current loop is
normal. Code that treats ok == 0 as "frozen" stops acting on alarms exactly
when the process gets noisy. Details and the exact bit list are in
docs/error-codes.md.
On SW 4.x the instrument reports an entrained-gas index (registers 1104 and 1113) alongside the density value. It is derived from frame-to-frame variance rather than absolute echo amplitude — which is the point, because amplitude alone cannot separate "more bubbles" from "more solids": both push it the same way. Alarm on the smoothed value (1113); the instantaneous one (1104) spikes on a single bubble crossing the beam.
This repository covers read-only registers — function code 0x04, the process values and diagnostics a PLC, DCS or historian needs.
Configuration, calibration and control writes are covered by the official Modbus manual supplied with the instrument. If you need it, ask us.
Register numbers are stable from SW 3.3.26 onward. The measurement block is append-only: newer firmware adds fields at the end and never moves an existing one, so a host written against an older map keeps working.
python tools/gen_tables.pyEdit the JSON, never the generated Markdown or CSV.
Pisonics builds inline density and concentration meters for mining, power generation, chemical processing and dredging. The PS7000 series measures slurry density by acoustic impedance, covering DN50 to DN1000 without a radioactive source.
- Product page — https://www.pisonics.com/products/PS7000
- How acoustic impedance measurement works — https://www.pisonics.com/guides/acoustic-impedance-explained
- Choosing a density meter — https://www.pisonics.com/guides/how-to-choose-density-meter
Questions about integration: info@pisonics.com
Documentation and register data: CC BY 4.0. Example code: MIT.