When the high voltage is turned off (e.g., during depolarization or an instrument reboot), detector calibration is affected. Consequently, a calibration file recorded immediately after a high-voltage shutdown should not be used to calibrate flare data taken, for instance, the previous day.
The current implementation of the routine for automatic download of the calibration file does not take into account high-voltage turnoff.
When the high voltage is turned off (e.g., during depolarization or an instrument reboot), detector calibration is affected. Consequently, a calibration file recorded immediately after a high-voltage shutdown should not be used to calibrate flare data taken, for instance, the previous day.
The current implementation of the routine for automatic download of the calibration file does not take into account high-voltage turnoff.