Current behavior
After the user performs a calibration of the pH probe, the Tank Controller checks the new slopes. If at least one of the slopes is not between 95% and 105% of ideal, then the Tank Controller warns the user that this appears to be a bad calibration. The user has two options:
- Override the warning, accepting the new calibration
- Reset to the probe's default ideal calibration
Desired behavior
Replace option 2 so that the user has two options:
- Override the warning, accepting the new calibration
- Cancel the new calibration, restoring the previous calibration